Abstract
We present an application of in vivo anterior segment imaging of the human eye with an ultrahigh speed swept source OCT instrument. For this purpose, a dedicated OCT system was designed and constructed. This instrument enables axial zooming by automatic reconfiguration of spectral sweep range; an enhanced imaging range mode enables imaging of the entire anterior segment while a high axial resolution mode provides detailed morphological information of the chamber angle and the cornea. The speed of 200,000 lines/s enables high sampling density in three-dimensional imaging of the entire cornea in 250 ms promising future applications of OCT for optical corneal topography, pachymetry and elevation maps. The results of a preliminary quantitative corneal analysis based on OCT data free form motion artifacts are presented. Additionally, a volumetric and real time reconstruction of dynamic processes, like pupillary reaction to light stimulus or blink-induced contact lens movements are demonstrated
Original language | English |
---|---|
Journal | Opt. Express |
Volume | 17 |
Issue number | 17 |
Pages (from-to) | 14880-14894 |
Number of pages | 15 |
DOIs | |