Skip to main navigation Skip to search Skip to main content

Abstract

This work evaluates the system performance of a dynamic cardiac SPECT (DC-SPECT) system using Monte Carlo simulations. We assessed the system’s sensitivity across the field of view (FOV) and evaluated spatial resolution using reconstructed images of a Derenzo phantom, a Jaszczak phantom, and an XCAT cardiac phantom. Results show that the system achieves a resolution of 5.5 mm for hot rods without background, and 6.0 mm for cold rods with warm background. The sensitivity ranges from 5 to 9 times greater than conventional dual-head SPECT over a 15 cm diameter spherical FOV. This study demonstrates that the current design of DC-SPECT has the potential to support effective cardiac imaging.

Conference

Conference2025 IEEE Nuclear Science Symposium (NSS), Medical Imaging Conference (MIC) and Room Temperature Semiconductor Detector Conference (RTSD)
Abbreviated title2025 IEEE NSS MIC RTSD
Country/TerritoryJapan
CityYokohama
Period01.11.2508.11.25