N2 - While previous generations of insertable cardiac monitors (ICMs) required a bedside monitor for remote monitoring (RM), the Confirm Rx™ ICM (Abbott, Chicago, IL, USA) utilizes Bluetooth®, Wi-Fi/cellular technology, and a smart device to connect to the RM system. We aimed to characterize compliance, connectivity, and event transmission timing with the Confirm Rx™ ICM RM system. The study cohort included American patients who received the Confirm Rx™ ICM with SharpSense™ technology within three months of release (May-July 2019). Compliance with RM was quantified as the proportion of patients registering the patient app on their smart device and transmitting at least once. Connectivity was measured as the median number of days between consecutive transmissions per patient. Event transmission time was measured from episode detection to availability on the Merlin.net™ RM system (Abbott). Time from transmission until review by a clinician was examined. Values for device connectivity, episode transmission timing, and clinician view times were reported as median [first quartile, third quartile]. Of 5,666 patients who received a Confirm Rx™ ICM, 97% registered their patient app and 92% transmitted data at least once. Among those utilizing RM (aged 66 ± 15 years; 49% female), connectivity occurred every 1.5 [1.2, 2.4] days, or 4.7 times per week. Patient-reported symptoms were transmitted to Merlin.net™ within 2.9 [2.1, 3.8] minutes of event onset and viewed by the clinician within 0.9 [0.4, 3.1] days, while device-detected episodes without symptoms were transmitted within 18.5 [11.2, 36.5] hours and then viewed within 0.8 [0.3, 2.5] days. This real-world study demonstrated excellent patient compliance with the smartphone-based RM paradigm enabled by Confirm Rx™, suggesting the suitability of this technology for future cardiac implantable devices.
AB - While previous generations of insertable cardiac monitors (ICMs) required a bedside monitor for remote monitoring (RM), the Confirm Rx™ ICM (Abbott, Chicago, IL, USA) utilizes Bluetooth®, Wi-Fi/cellular technology, and a smart device to connect to the RM system. We aimed to characterize compliance, connectivity, and event transmission timing with the Confirm Rx™ ICM RM system. The study cohort included American patients who received the Confirm Rx™ ICM with SharpSense™ technology within three months of release (May-July 2019). Compliance with RM was quantified as the proportion of patients registering the patient app on their smart device and transmitting at least once. Connectivity was measured as the median number of days between consecutive transmissions per patient. Event transmission time was measured from episode detection to availability on the Merlin.net™ RM system (Abbott). Time from transmission until review by a clinician was examined. Values for device connectivity, episode transmission timing, and clinician view times were reported as median [first quartile, third quartile]. Of 5,666 patients who received a Confirm Rx™ ICM, 97% registered their patient app and 92% transmitted data at least once. Among those utilizing RM (aged 66 ± 15 years; 49% female), connectivity occurred every 1.5 [1.2, 2.4] days, or 4.7 times per week. Patient-reported symptoms were transmitted to Merlin.net™ within 2.9 [2.1, 3.8] minutes of event onset and viewed by the clinician within 0.9 [0.4, 3.1] days, while device-detected episodes without symptoms were transmitted within 18.5 [11.2, 36.5] hours and then viewed within 0.8 [0.3, 2.5] days. This real-world study demonstrated excellent patient compliance with the smartphone-based RM paradigm enabled by Confirm Rx™, suggesting the suitability of this technology for future cardiac implantable devices.
UR - http://www.scopus.com/inward/record.url?scp=85111998989&partnerID=8YFLogxK
U2 - 10.19102/icrm.2021.120806
DO - 10.19102/icrm.2021.120806
M3 - Journal articles
AN - SCOPUS:85111998989
SN - 2156-3977
VL - 12
SP - 4613
EP - 4620
JO - Journal of Innovations in Cardiac Rhythm Management
JF - Journal of Innovations in Cardiac Rhythm Management
IS - 7
ER -