Abstract
Magnetic Particle Imaging (MPI) has been introduced as a modality that allows for the acquisition of three-dimensional functional images with high sensitivity in real time. Here, alternative coil topologies are presented that differ significantly from the original set-up. Two novel coil topologies will be presented. Beside an asymmetric coil topology, where all field generating coils are arranged on a single side, an effective coil assembly has been accomplished that creates a field-free line for spatial coding. The alternative coil topologies may overcome the problem of a confined measurement field or lead to an increase of the sensitivity of MPI.
Original language | English |
---|---|
Title of host publication | Medical Imaging 2011: Biomedical Applications in Molecular, Structural, and Functional Imaging |
Number of pages | 6 |
Volume | 7965 |
Publisher | SPIE |
Publication date | 16.05.2011 |
Article number | 79650T |
ISBN (Print) | 978-081948507-6 |
DOIs | |
Publication status | Published - 16.05.2011 |