Skip to main navigation Skip to search Skip to main content

Influence of System Matrix Concentration and Frequency Selection on Reconstructed MPI Signals

Sathurya Jegatheeswaran*, Hanna Meenen, Thorsten M. Buzug, Ksenija Gräfe

*Corresponding author for this work

Abstract

In Magnetic Particle Imaging (MPI), the system matrix (SM) characterizes the spatial encoding properties of the imaging system and the magnetic nanoparticle response. Ideally, for linear particle magnetization behavior, the concentration at which the SM is measured should not affect the reconstructed image. In this study, two SMs recorded with tracer concentrations of 8.5 mg(Fe)/ml and 1.92 mg(Fe)/ml were investigated. We show that, although the same SM is used, reconstruction discrepancies occur when the frequency coverage differs. We therefore propose to determine the usable frequency components based on the lower concentration SM to ensure a consistent reconstruction basis.

Original languageEnglish
Article number2603026
JournalInternational Journal on Magnetic Particle Imaging
Volume12
Issue number1
ISSN2365-9033
DOIs
Publication statusPublished - 2026

Funding

FundersFunder number
Damp Stiftung2022-08
European Cooperation in Science and TechnologyCA23132

    Fingerprint

    Dive into the research topics of 'Influence of System Matrix Concentration and Frequency Selection on Reconstructed MPI Signals'. Together they form a unique fingerprint.