Abstract

We present full volumetric high speed OCT imaging of the retina with multiple settings varying in volume size and volume rate. The volume size ranges from 255x255 A-scans to 160x40 A-scans with 450 samples per depth scan with volume rates varying between 20.8 V/s for the largest volumes to 195.2 V/s for the smallest. The system is based on a 1060nm Fourier domain mode locked (FDML) laser with 1.6MHz line rate. Scanning along the fast axis is performed with a 2.7 kHz or 4.3 kHz resonant scanner operated in bidirectional scanning mode, while a standard galvo scanner is used for the slow axis. The performance is analyzed with respect to various potential applications, like intraoperative OCT.
Original languageEnglish
Title of host publicationOptical Coherence Imaging Techniques and Imaging in Scattering Media
EditorsBrett E. Bouma, Maciej Wojtkowski
Number of pages8
Volume95410Z
PublisherSPIE
Publication date22.07.2015
Pages95410Z-95410Z-8
ISBN (Print)9781628417067
DOIs
Publication statusPublished - 22.07.2015
EventEUROPEAN CONFERENCES ON BIOMEDICAL OPTICS - München, Germany
Duration: 21.06.201525.06.2015
Conference number: 131538
https://spie.org/conferences-and-exhibitions?webSyncID=8ea9e3e7-7b6b-4836-788d-177c8547e3b2&sessionGUID=7a4397d1-7ae8-5844-19e9-c62bcc516bb5&_ga=2.220047066.216408110.1526970521-1856364488.1526970521&SSO=1

Fingerprint

Dive into the research topics of 'High definition in vivo retinal volumetric video rate OCT at 0.6 Giga-voxels per second'. Together they form a unique fingerprint.