Abstract
OBJECTIVES: Intra-arterial digital subtraction angiography (IA-DSA), an invasive procedure, is the current reference examination after percutaneous transluminal angioplasty and stenting for the detection of in-stent restenosis (ISR). In this phantom study, we evaluated flat-panel angiographic computed tomography after intravenous contrast agent application (IV-ACT) and multidetector computed tomographic angiography (MDCTA) as potential noninvasive follow-up alternatives after intracranial percutaneous transluminal angioplasty and stenting.
MATERIALS AND METHODS: We simulated an intracranial vessel using a silicon tube placed inside a human skull. Three different stent systems were deployed inside the silicon tubes, each with diameters of 3 or 4 mm. Three grades of ISR (25%, 50%, and 75%) were simulated. The IA-DSA and IV-ACT examinations were performed on a flat-panel detector angiography system. The MDCTA images were acquired with a 128-slice computed tomographic scanner. The mean stenosis diameters, measured with each technique, were compared using the Bland-Altman plot. The difference between the known stenosis diameter and the measured stenosis diameter was calculated for each examination.
RESULTS: Stenosis measurements on the IA-DSA images showed no statistically significant differences compared with the known stenosis diameters (P = 0.19). In the 3-mm stent category, when compared with the known stenosis diameter, mean (SD) differences of 0.01 (0.15) mm, 0.03 (0.24) mm and 0.16 (0.5) mm were calculated for the IA-DSA, IV-ACT, and MDCTA stenosis measurements, respectively. As for the 4-mm stents, IA-DSA and IV-ACT were again very accurate, with mean (SD) differences of -0.03 (0.11) mm and 0.07 (0.19) mm, respectively, compared with the known stenosis diameters, whereas MDCTA overestimated ISR, with a mean (SD) difference of 0.49 (0.53) mm. The Bland-Altman plots show a mean (SD) difference of 0.08 (0.2) mm between IA-DSA and IV-ACT (95% confidence interval, 0.05-0.11) and a mean (SD) difference of 0.34 (0.56) mm between IA-DSA and MDCTA measurements (95% confidence interval, 0.25-0.42).
CONCLUSIONS: In our phantom study, IA-DSA was the only examination to predict accurately degrees of stenosis compared with the known stenosis diameters. The results of the IV-ACT measurements were comparable with those of IA-DSA. Multidetector computed tomographic angiography was less accurate in the quantification of stenosis, usually overestimating ISR.
| Original language | English |
|---|---|
| Journal | Investigative Radiology |
| Volume | 48 |
| Issue number | 2 |
| Pages (from-to) | 98-103 |
| Number of pages | 6 |
| ISSN | 0020-9996 |
| DOIs | |
| Publication status | Published - 02.2013 |
UN SDGs
This output contributes to the following UN Sustainable Development Goals (SDGs)
-
SDG 3 Good Health and Well-being
Fingerprint
Dive into the research topics of 'Evaluation of noninvasive follow-up methods for the detection of intracranial in-stent restenosis: a phantom study'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver