Projects per year
Abstract
This work presents an evaluation study using a force feedback evaluation framework for a novel direct needle force volume rendering concept in the context of liver puncture simulation. PTC/PTCD puncture interventions targeting the bile ducts have been selected to illustrate this concept. The haptic algorithms of the simulator system are based on (1) partially segmented patient image data and (2) a non-linear spring model effective at organ borders. The primary aim is to quantitatively evaluate force errors caused by our patient modeling approach, in comparison to haptic force output obtained from using gold-standard, completely manually-segmented data. The evaluation of the force algorithms compared to a force output from fully manually segmented gold-standard patient models, yields a low mean of 0.12 N root mean squared force error and up to 1.6 N for systematic maximum absolute errors. Force errors were evaluated on 31,222 preplanned test paths from 10 patients. Only twelve percent of the emitted forces along these paths were affected by errors. This is the first study evaluating haptic algorithms with deformable virtual patients in silico. We prove haptic rendering plausibility on a very high number of test paths. Important errors are below just noticeable differences for the hand-Arm system.
Original language | English |
---|---|
Article number | 671 |
Journal | Scientific Reports |
Volume | 7 |
Issue number | 1 |
ISSN | 2045-2322 |
DOIs | |
Publication status | Published - 01.12.2017 |
Fingerprint
Dive into the research topics of 'Evaluation of Direct Haptic 4D Volume Rendering of Partially Segmented Data for Liver Puncture Simulation'. Together they form a unique fingerprint.Projects
- 1 Finished
-
Patient-specific 4D Virtual Reality Simulation of Punctures and Radiofrequency Ablations in Breathing Virtual Body Models
Handels, H. (Principal Investigator (PI)) & Mastmeyer, A. (Project Staff)
01.04.10 → 31.03.18
Project: DFG Projects › DFG Individual Projects