Continuous measurement of cardiac output during aortic crossclamping by the oesophageal Doppler monitor ODM 1

K. F. Klotz*, S. Klingsiek, M. Singer, H. Wenk, S. Eleftheriadis, H. Kuppe, P. Schmucker

*Corresponding author for this work
45 Citations (Scopus)

Abstract

We have compared the Doppler against the thermodilution technique for measurement of cardiac output in six patients during aortic surgery. The correlation coefficient between the two methods was between 0.76 and 0.84 during the different periods of the operation. Using the integral nomogram instead of direct calibration, the Doppler system underestimated cardiac output in atherosclerotic patients. However, the Doppler method did register accurately significant changes in cardiac output.

Original languageEnglish
JournalBritish Journal of Anaesthesia
Volume74
Issue number6
Pages (from-to)655-660
Number of pages6