Automated flow quantification in valvular heart disease based on backscattered Doppler power analysis: Implementation on matrix-array ultrasound imaging systems

Thomas Buck*, Shawn M. Hwang, Björn Plicht, Ronald A. Mucci, Peter Hunold, Raimund Erbel, Robert A. Levine

*Corresponding author for this work
5 Citations (Scopus)

Abstract

Objective: Cardiac ultrasound imaging systems are limited in the noninvasive quantification of valvular regurgitation due to indirect measurements and inaccurate hemodynamic assumptions. We recently demonstrated that the principle of integration of backscattered acoustic Doppler power times velocity can be used for flow quantification in valvular regurgitation directly at the vena contracta of a regurgitant flow jet. We now aimed to accomplish implementation of automated Doppler power flow analysis software on a standard cardiac ultrasound system utilizing novel matrix-array transducer technology with detailed description of system requirements, components and software contributing to the system. Methods: This system based on a 3.5 MHz, matrix-array cardiac ultrasound scanner (Sonos 5500, Philips Medical Systems) was validated by means of comprehensive experimental signal generator trials, in vitro flow phantom trials and in vivo testing in 48 patients with mitral regurgitation of different severity and etiology using magnetic resonance imaging (MRI) for reference. Results: All measurements displayed good correlation to the reference values, indicating successful implementation of automated Doppler power flow analysis on a matrix-array ultrasound imaging system. Systematic underestimation of effective regurgitant orifice areas >0.65 cm2 and volumes >40 ml was found due to currently limited Doppler beam width that could be readily overcome by the use of new generation 2D matrix-array technology. Conclusion: Automated flow quantification in valvular heart disease based on backscattered Doppler power can be fully implemented on board a routinely used matrix-array ultrasound imaging systems. Such automated Doppler power flow analysis of valvular regurgitant flow directly, noninvasively, and user independent overcomes the practical limitations of current techniques.

Original languageEnglish
JournalInternational Journal of Cardiovascular Imaging
Volume24
Issue number5
Pages (from-to)463-477
Number of pages15
ISSN1569-5794
DOIs
Publication statusPublished - 01.06.2008

Fingerprint

Dive into the research topics of 'Automated flow quantification in valvular heart disease based on backscattered Doppler power analysis: Implementation on matrix-array ultrasound imaging systems'. Together they form a unique fingerprint.

Cite this