Artifact free reconstruction with the system matrix approach by overscanning the field-free-point trajectory in magnetic particle imaging

A. Weber, F. Werner, J. Weizenecker, T. M. Buzug, T. Knopp

5 Citations (Scopus)

Abstract

Magnetic particle imaging is a tracer-based imaging method that utilizes the non-linear magnetization response of iron-oxide for determining their spatial distribution. The method is based on a sampling scheme where a sensitive spot is moved along a trajectory that captured a predefined field-of-view (FOV). However, particles outside the FOV also contribute to the measurement signal due to their rotation and the non-sharpness of the sensitive spot. In the present work we investigate artifacts that are induced by particles not covered by the FOV and show that the artifacts can be mitigated by using a system matrix that covers not only the region of interest but also a certain area around the FOV. The findings are especially relevant when using a multi-patch acquisition scheme where the boundaries of neighboring patches have to be handled.

Original languageEnglish
JournalPhysics in Medicine and Biology
Volume61
Issue number2
Pages (from-to)475-487
Number of pages13
ISSN0031-9155
DOIs
Publication statusPublished - 18.12.2015

Fingerprint

Dive into the research topics of 'Artifact free reconstruction with the system matrix approach by overscanning the field-free-point trajectory in magnetic particle imaging'. Together they form a unique fingerprint.

Cite this