Skip to main navigation Skip to search Skip to main content

An evolutionary approach to continuously estimate CPR quality parameters from a wrist-worn inertial sensor

Christian Lins*, Björn Friedrich, Andreas Hein, Sebastian Fudickar

*Corresponding author for this work

Abstract

Cardiopulmonary resuscitation (CPR) is one of the most critical emergency interventions for sudden cardiac arrest. In this paper, a robust sinusoidal model-fitting method based on a Evolution Strategy inspired algorithm for CPR quality parameters – naming chest compression frequency and depth – as measured by an inertial measurement unit (IMU) attached to the wrist is presented. The proposed approach will allow bystanders to improve CPR as part of a continuous closed-loop support system once integrated into a smartphone or smartwatch application. By evaluating the model’s precision with data recorded by a training mannequin as reference standard, a variance for the compression frequency of ± 2.22 compressions per minute (cpm) has been found for the IMU attached to the wrist. It was found that this previously unconsidered position and thus, the use of smartwatches is a suitable alternative to the typical placement of phones in hand for CPR training.

Original languageEnglish
JournalHealth and Technology
Volume12
Issue number1
Pages (from-to)161-173
Number of pages13
ISSN2190-7188
DOIs
Publication statusPublished - 01.2022

Funding

FundersFunder number
Niedersächsische Ministerium für Wissenschaft und Kultur

    UN SDGs

    This output contributes to the following UN Sustainable Development Goals (SDGs)

    1. SDG 3 - Good Health and Well-being
      SDG 3 Good Health and Well-being
    2. SDG 9 - Industry, Innovation, and Infrastructure
      SDG 9 Industry, Innovation, and Infrastructure

    Fingerprint

    Dive into the research topics of 'An evolutionary approach to continuously estimate CPR quality parameters from a wrist-worn inertial sensor'. Together they form a unique fingerprint.

    Cite this