A high power driving and selection field coil for an open MPI scanner

G. Bringout, H. Wojtczyk, W. Tenner, M. Graeser, M. Grüttner, J. Hägele, R. Duschka, N. Panagiotopoulos, F.M. Vogt, J. Barkhausen, T.M. Buzug

Abstract

Three Litz wires have been used in parallel in order to decrease the layer to layer voltage down to 4 kV and the coil voltage down to 8 kV. A two millimeter thick epoxy plate has been glued between each layer, in order to prevent any electrical breakdown between the layers. A special epoxy has been used to minimize the inner temperature of the coil, which reached 180°C with a dissipated power of 5.2 kW. The use of parallel Litz wire leads to currents having different phase angle in every wire which has to be corrected.
Original languageEnglish
Title of host publication 2013 International Workshop on Magnetic Particle Imaging (IWMPI)
Number of pages1
PublisherIEEE
Publication date01.03.2013
Pages1-1
ISBN (Print)978-1-4673-5520-9
ISBN (Electronic)978-1-4673-5522-3
DOIs
Publication statusPublished - 01.03.2013
Event3th International Workshop on Magnetic Particle Imaging (IWMPI 2013) - Berkeley, United States
Duration: 23.03.201324.03.2013

Fingerprint

Dive into the research topics of 'A high power driving and selection field coil for an open MPI scanner'. Together they form a unique fingerprint.