Skip to main navigation Skip to search Skip to main content

3D Druck des NEMA NU 4-2008 Phantoms für die Kleintier-PET-Bildgebung – eine Vergleichsstudie mit verschiedenen Bildgebungsmodalitäten

M. Grehn, J. Kurth, M. Rafecas, Jan Stenzel

Abstract

The study examined the imaging and material properties of two 3D printed versions of the NEMA NU 4-2008 image quality phantom for small animal Positron Emission Tomography compared with a manufactured and a commercial one (used as reference). Based on the present results it could be shown that the support material of the 3D printer has an influence on the later phantom quality. Also of importance is the alignment of the print layers. A basic suitability of 3D printed phantoms for all imaging modalities does not seem to be admissible and should be carefully evaluated in advance.
Original languageGerman
Title of host publication47. Jahrestagung der DGMP und 19. Jahrestagung der DS-ISMRM
EditorsL. M. Schreiber, J.-B. Hövener
Number of pages2
PublisherDGMP
Publication date2016
Pages125-126
ISBN (Print)978-3-9816508-2-2
Publication statusPublished - 2016
Event47. Jahrestagung der Deutschen Gesellschaft für Medizinische Physik (DGMP) e. V.
19. Jahrestagung der Deutschen Sektion der International Society for Magnetic Resonance in Medicine (ISMRM) e. V.
- Würzburg, Germany
Duration: 07.09.2016 → 10.09.2016

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being
  2. SDG 9 - Industry, Innovation, and Infrastructure
    SDG 9 Industry, Innovation, and Infrastructure

Research Areas and Centers

  • Academic Focus: Biomedical Engineering

DFG Research Classification Scheme

  • 2.22-32 Medical Physics, Biomedical Technology

Cite this