Abstract
We present an 840nm Fourier domain mode locked (FDML) based swept-source OCT system with an A-Scan rate of 1.7MHz. The laser shows promising performance for retinal imaging with 10mW output power and a theoretical axial resolution in water of 12μm. The system demonstrates shot-noise-limited behavior, achieving a sensitivity of 84dB and no roll-off over the analog bandwidth of the balanced photodetector. In-vivo ophthalmic images are presented to show the performances of the laser.
| Original language | English |
|---|---|
| Title of host publication | Ophthalmic Technologies XXXV |
| Editors | Daniel X. Hammer, Derek Nankivil, Yuankai K. Tao |
| Number of pages | 1 |
| Volume | 13300 |
| Publisher | SPIE |
| Publication date | 03.2025 |
| Pages | 1330004 |
| DOIs | |
| Publication status | Published - 03.2025 |
UN SDGs
This output contributes to the following UN Sustainable Development Goals (SDGs)
-
SDG 9 Industry, Innovation, and Infrastructure
Research Areas and Centers
- Academic Focus: Biomedical Engineering
DFG Research Classification Scheme
- 2.22-32 Medical Physics, Biomedical Technology