Ultrahigh-speed intravascular optical coherence tomography imaging at 3200 frames per second

Tianshi Wang, Wolfgang Wieser, Geert Springeling, Robert Beurskens, Charles T Lancee, Tom Pfeiffer, Antonius F W van der Steen, Robert Huber, Gijs van Soest

Abstract

We demonstrated intravascular OCT imaging with frame rate up to 3.2 kHz (192,000 rpm scanning). This was achieved by using a custom-built catheter in which the circumferential scanning was actuated by a 1.0 mm diameter synchronous motor. The OCT system was based on a Fourier Domain Mode Locked laser operating at an A-line rate of 1.6 MHz. The diameter of the catheter was 1.1 mm at the tip. Ex vivo images of human coronary artery (~78.4 mm length) were acquired at a pullback speed of 100 mm/s. True 3D volumetric imaging of the entire artery, with adequate sampling in all dimensions, was performed in < 1 second acquisition time.
OriginalspracheEnglisch
TitelOptical Coherence Tomography and Coherence Techniques VI
Redakteure/-innenBrett E Bouma, Rainer A Leitgeb
Seitenumfang4
Band88020O
Herausgeber (Verlag)SPIE
Erscheinungsdatum18.06.2013
Seiten8802 - 8802 - 4
ISBN (Print)9780819496515
DOIs
PublikationsstatusVeröffentlicht - 18.06.2013
VeranstaltungEUROPEAN CONFERENCES ON BIOMEDICAL OPTICS - Munich, Deutschland
Dauer: 12.03.201316.03.2013

Fingerprint

Untersuchen Sie die Forschungsthemen von „Ultrahigh-speed intravascular optical coherence tomography imaging at 3200 frames per second“. Zusammen bilden sie einen einzigartigen Fingerprint.