Towards Multi-Directional OCT for Speckle Noise Reduction

Lukas Ramrath, Guillermo Moreno, Heike Hedwig Müller, Tim Bonin, Gereon Hüttmann, Achim Schweikard

Abstract

Multi-directional optical coherence tomography (MD-OCT) applies and extends the concept of angular compounding for speckle noise reduction to the area of OCT imaging. OCT images are acquired from a wide range of angles of view. Averaging of the rotated images therefore requires compensation of the parallax which is achieved by simple image registration for image reconstruction. Test measurements of a sample structure in a low and highly scattering environment show that the method improves the signal-to-noise ratio by a factor of 4 and hence reduces speckle noise significantly. Experimental results also show that the proposed averaging increases the performance of common edge-detection algorithms.

OriginalspracheEnglisch
TitelMICCAI 2008: Medical Image Computing and Computer-Assisted Intervention – MICCAI 2008
Seitenumfang9
Band5241 LNCS
Herausgeber (Verlag)Springer Verlag
Erscheinungsdatum01.12.2008
Seiten815-823
ISBN (Print)978-3-540-85987-1
ISBN (elektronisch)978-3-540-85988-8
DOIs
PublikationsstatusVeröffentlicht - 01.12.2008
Veranstaltung11th International Conference on Medical Image Computing and Computer-Assisted Intervention - New York, USA / Vereinigte Staaten
Dauer: 06.09.200810.09.2008
Konferenznummer: 73795

UN SDGs

Dieser Output leistet einen Beitrag zu folgendem(n) Ziel(en) für nachhaltige Entwicklung

  1. SDG 3 – Gesundheit und Wohlergehen
    SDG 3 – Gesundheit und Wohlergehen
  2. SDG 9 – Industrie, Innovation und Infrastruktur
    SDG 9 – Industrie, Innovation und Infrastruktur

Fingerprint

Untersuchen Sie die Forschungsthemen von „Towards Multi-Directional OCT for Speckle Noise Reduction“. Zusammen bilden sie einen einzigartigen Fingerprint.

Zitieren