Zur Hauptnavigation wechseln Zur Suche wechseln Zum Hauptinhalt wechseln

Time-to-effect guided pulmonary vein isolation utilizing the third-generation versus second generation cryoballoon: One year clinical success

Christian Hendrik Heeger*, Christopher Schuette, Valentina Seitelberger, Erik Wissner, Andreas Rillig, Shibu Mathew, Bruno Reissmann, Christine Lemes, Tilman Maurer, Thomas Fink, Osamu Inaba, Naotaka Hashiguchi, Francesco Santoro, Feifan Ouyang, Karl Heinz Kuck, Andreas Metzner

*Korrespondierende/r Autor/-in für diese Arbeit

Abstract

Background: The second-generation cryoballoon (CB2) provides effective and durable pulmonary vein isolation (PVI) associated with encouraging and reproducible clinical outcome data. The latest--generation cryoballoon (CB3) incorporates a 40% shorter distal tip, thus allowing for an increased rate of PVI real-time signal recording and facilitating individualized ablation strategies taking the time-to--effect (TTE) into account. However, whether this characteristic translates into favorable clinical success has not been evaluated yet. Herein was investigated 1-year clinical success after CB3 in comparison to CB2 based-PVI. Methods: One hundred and ten consecutive patients with paroxysmal or short-standing persistent atrial fibrillation (AF) underwent CB2 (n = 55 patients)-or CB3 (n = 55 patients)-based PVI. The freeze-cycle duration was set to TTE + 120 s if TTE could be recorded, otherwise a fixed freeze-cycle duration of 180 s was applied. Results: A total of 217/218 (99%, CB3) and 217/217 (100%, CB2) pulmonary veins (PV) were successfully isolated. The real-time PVI visualization rate was 69.2% (CB3) and 54.8% (CB2; p = 0.0392). The mean freeze-cycle duration was 194 ± 77 s (CB3) and 206 ± 85 s (CB2; p = 0.132), respectively. During a median follow-up of 409 days (interquartile range [IQR] 378–421, CB3) and 432 days (IQR 394–455, CB2) 73.6% (CB3) and 73.1% of patients (CB2) remained in stable sinus rhythm after a single procedure (p = 0.806). Conclusions: A higher rate of real-time electrical PV recordings was seen using the CB3 as compared to CB2. There was no difference in 1-year clinical follow-up.

OriginalspracheEnglisch
ZeitschriftCardiology Journal
Jahrgang26
Ausgabenummer4
Seiten (von - bis)368-374
Seitenumfang7
ISSN1897-5593
DOIs
PublikationsstatusVeröffentlicht - 05.06.2019

Fördermittel

received travel grants by St. Jude Medical, Bio-tronik and Medtronic. Erik Wissner is a consultant to Medtronic. Karl-Heinz Kuck received travel grants and research grants from Biosense Webster, Stereotaxis, Prorhythm, Medtronic, Edwards, Cryocath, and is a consultant to St. Jude Medical, Biosense Webster, Prorhythm, and Stereotaxis. He received speaker’s honoraria from Medtronic. Andreas Metzner received speaker’s honoraria and travel grants from Medtronic, Biosense Webster

UN SDGs

Dieser Output leistet einen Beitrag zu folgendem(n) Ziel(en) für nachhaltige Entwicklung

  1. SDG 3 – Gesundheit und Wohlergehen
    SDG 3 – Gesundheit und Wohlergehen

Fingerprint

Untersuchen Sie die Forschungsthemen von „Time-to-effect guided pulmonary vein isolation utilizing the third-generation versus second generation cryoballoon: One year clinical success“. Zusammen bilden sie einen einzigartigen Fingerprint.

Zitieren