Zur Hauptnavigation wechseln Zur Suche wechseln Zum Hauptinhalt wechseln

Three-dimensional anisotropic regularization for limited angle tomography

C. F. Isernhagen, D. Schäfer, M. Grass, T. M. Buzug

Abstract

Limited angle tomography is a challenging task in medical imaging. Due to practical limitations during the image acquisition, the sinogram is recorded incompletely and thus the quality of the reconstruction is deteriorated by streak artifacts. These artifacts are characterized by fast changes of the local intensity gradients and increase the total variation (TV). Generally, an energy functional is optimized which leads to a minimized Total Variation Minimization (TVM). As an outcome, noise and artifacts are reduced while edges are preserved. Anyway, often the orientation of the streak artifacts is not considered at all. Therefore, anisotropic regularization is used to reduce noise and distortions under specific directions.

OriginalspracheEnglisch
ZeitschriftCurrent Directions in Biomedical Engineering
Jahrgang1
Ausgabenummer1
Seiten (von - bis)283-285
Seitenumfang3
DOIs
PublikationsstatusVeröffentlicht - 2015

UN SDGs

Dieser Output leistet einen Beitrag zu folgendem(n) Ziel(en) für nachhaltige Entwicklung

  1. SDG 3 – Gesundheit und Wohlergehen
    SDG 3 – Gesundheit und Wohlergehen
  2. SDG 9 – Industrie, Innovation und Infrastruktur
    SDG 9 – Industrie, Innovation und Infrastruktur

Fingerprint

Untersuchen Sie die Forschungsthemen von „Three-dimensional anisotropic regularization for limited angle tomography“. Zusammen bilden sie einen einzigartigen Fingerprint.

Zitieren