Zur Hauptnavigation wechseln Zur Suche wechseln Zum Hauptinhalt wechseln

SNR and Discretization Enhancement for System Matrix Determination by Decreasing the Gradient in Magnetic Particle Imaging

Matthias Graeser, Anselm von Gladiß, Thomas Friedrich, Thorsten M. Buzug

Abstract

In system matrix (SM) based reconstruction, the physical resolution is often within the range of the SM discretization. This is caused by the signal to noise ratio (SNR) decrease following a discretization increase due to the smaller particle sample volume. As the SNR affects the resolution of the image as well, it is necessary to decouple the SNR and discretization. In this work, a calibration protocol is presented which enhances either the SNR or discretization by reducing the gradient strength within the system calibration. This new protocol results in higher resolution and better image quality.
OriginalspracheEnglisch
Aufsatznummer1703019
ZeitschriftInternational Journal on Magnetic Particle Imaging
Jahrgang3
Ausgabenummer1
Seitenumfang4
DOIs
PublikationsstatusVeröffentlicht - 23.03.2017

UN SDGs

Dieser Output leistet einen Beitrag zu folgendem(n) Ziel(en) für nachhaltige Entwicklung

  1. SDG 3 – Gesundheit und Wohlergehen
    SDG 3 – Gesundheit und Wohlergehen
  2. SDG 9 – Industrie, Innovation und Infrastruktur
    SDG 9 – Industrie, Innovation und Infrastruktur

Fingerprint

Untersuchen Sie die Forschungsthemen von „SNR and Discretization Enhancement for System Matrix Determination by Decreasing the Gradient in Magnetic Particle Imaging“. Zusammen bilden sie einen einzigartigen Fingerprint.

Zitieren