Abstract
Magnetic Particle Imaging (MPI) is a new imaging modality that allows to detect magnetizable nanoparticles. Recently, resolution and sensitivity of this method was investigated in a simulation study. Since the reconstruction problem in MPI is presumably ill-posed, Tikhonov regularization was applied to achieve optimal imaging quality for a certain set of acquisition parameters. The purpose of this work is to analyze the conditioning of the reconstruction problem in MPI by singular value decomposition. Considering the singular values, the ill-posedness of the reconstmction can be shown. For computing the optimal regularization parameter automatically, an algorithm is applied that exploits the L-curve.
| Originalsprache | Englisch |
|---|---|
| Titel | 2008 IEEE Nuclear Science Symposium Conference Record |
| Seitenumfang | 5 |
| Herausgeber (Verlag) | IEEE |
| Erscheinungsdatum | 01.12.2008 |
| Seiten | 4525-4529 |
| Aufsatznummer | 4774296 |
| ISBN (Print) | 978-1-4244-2714-7, 978-1-4244-2715-4 |
| DOIs | |
| Publikationsstatus | Veröffentlicht - 01.12.2008 |
| Veranstaltung | 2008 IEEE Nuclear Science Symposium Conference Record - Dresden, Deutschland Dauer: 19.10.2008 → 25.10.2008 Konferenznummer: 76975 |
UN SDGs
Dieser Output leistet einen Beitrag zu folgendem(n) Ziel(en) für nachhaltige Entwicklung
-
SDG 3 – Gesundheit und Wohlergehen
-
SDG 9 – Industrie, Innovation und Infrastruktur
Fingerprint
Untersuchen Sie die Forschungsthemen von „Singular value analysis for Magnetic Particle Imaging“. Zusammen bilden sie einen einzigartigen Fingerprint.Zitieren
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver