Singular value analysis for Magnetic Particle Imaging

Tobias Knopp, Sven Biederer, Timo Sattel, Thorsten M. Buzug

Abstract

Magnetic Particle Imaging (MPI) is a new imaging modality that allows to detect magnetizable nanoparticles. Recently, resolution and sensitivity of this method was investigated in a simulation study. Since the reconstruction problem in MPI is presumably ill-posed, Tikhonov regularization was applied to achieve optimal imaging quality for a certain set of acquisition parameters. The purpose of this work is to analyze the conditioning of the reconstruction problem in MPI by singular value decomposition. Considering the singular values, the ill-posedness of the reconstmction can be shown. For computing the optimal regularization parameter automatically, an algorithm is applied that exploits the L-curve.

OriginalspracheEnglisch
Titel2008 IEEE Nuclear Science Symposium Conference Record
Seitenumfang5
Herausgeber (Verlag)IEEE
Erscheinungsdatum01.12.2008
Seiten4525-4529
Aufsatznummer4774296
ISBN (Print)978-1-4244-2714-7, 978-1-4244-2715-4
DOIs
PublikationsstatusVeröffentlicht - 01.12.2008
Veranstaltung2008 IEEE Nuclear Science Symposium Conference Record - Dresden, Deutschland
Dauer: 19.10.200825.10.2008
Konferenznummer: 76975

Fingerprint

Untersuchen Sie die Forschungsthemen von „Singular value analysis for Magnetic Particle Imaging“. Zusammen bilden sie einen einzigartigen Fingerprint.

Zitieren