Safety Aspects for a Pre-clinical Magnetic Particle Imaging Scanner

Abstract

Magnetic Particle Imaging is a promising new imaging technique using magnetic fields to image magnetic tracer material in the body. As with MRI systems, time varying magnetic fields raise some safety issues. The stimulation of peripheral nerves and tissues is one of them. In the paper, the stimulation thresholds are explained and an evaluation of the stimulation generated by a pre-clinical scanner is calculated. It appears clearly that, even if driving fields of high amplitude are used, cardiac arrhythmias are unlikely to be induced. However, it is yet unclear whether some peripheral nerve stimulation may be induced.
OriginalspracheEnglisch
TitelMagnetic Particle Imaging
Redakteure/-innenThorsten M. Buzug, Jörn Borgert
Seitenumfang5
Band140
Herausgeber (Verlag)Springer Berlin Heidelberg
Erscheinungsdatum2012
Seiten355-359
ISBN (Print)978-3-642-24132-1
ISBN (elektronisch)978-3-642-24133-8
DOIs
PublikationsstatusVeröffentlicht - 2012
Veranstaltung2th International Workshop on Magnetic Particle Imaging (IWMPI 2012) - Lübeck, Deutschland
Dauer: 15.03.2012 → 16.03.2012
https://www.iwmpi.org/home/previous-conferences/2012/

UN SDGs

Dieser Output leistet einen Beitrag zu folgendem(n) Ziel(en) für nachhaltige Entwicklung

  1. SDG 3 – Gesundheit und Wohlergehen
    SDG 3 – Gesundheit und Wohlergehen
  2. SDG 9 – Industrie, Innovation und Infrastruktur
    SDG 9 – Industrie, Innovation und Infrastruktur

Fingerprint

Untersuchen Sie die Forschungsthemen von „Safety Aspects for a Pre-clinical Magnetic Particle Imaging Scanner“. Zusammen bilden sie einen einzigartigen Fingerprint.

Dieses zitieren