Safe and reliable sound threshold measures with direct vibration of the ossicular chain

Marcus M. Maassen*, Jesus Rodriguez Jorge, Stephan Herberhold, Reinhard Vonthein, Rainer Zimmermann, Ingo Baumann, Sibylle Brosch, Paul S. Mauz, Markus Pfister, Ilse M. Zalaman, Hubert Löwenheim, Hans P. Zenner

*Korrespondierende/r Autor/-in für diese Arbeit

Abstract

Objectives: The aim of the study was to evaluate the safety and feasibility of piezoelectric malleus vibration audiometer (MVA), which presents micromechanical vibrations to the umbo membranae tympani. Study Design: Phase I study performed in a tertiary referral center (University Hospital). Methods: The coupling rod of the MVA was moved slowly through the outer ear canal toward the eardrum with a micromanipulator. Coupling was completed when the rod tip touched the umbo membranae tympani. Basic audiologic measures of sound threshold obtained with direct stimulation of the malleus are presented. We used MANOVA (multivariate repeated measures ANOVA) to investigate the repeatability of MVA thresholds from one day to the other and when decoupling and retracting the coupling rod 2 mm off the umbo. We also selected the MANOVA to test for unwanted bone-conduction threshold shifts after MVA application. We assessed normality of the data by quantile-quantile plots of the residuals. Results: Twenty-eight male and 10 female subjects with normal hearing, 22.2 to 34.6 years old (median age, 27.2 yr) underwent an examination. Thirty-six subjects underwent MVA, because 2 of the 38 subjects who volunteered for the study have not undergone the procedure due to the external auditory canal anatomy preventing application of the MVA. The results show that it is possible to safely and reliably measure thresholds of direct vibration of the ossicular chain. Using pure tone audio grams, no pure tone bone- and/or air-conduction threshold shifts occurred after the procedure. None of the subjects reported any other ear-related symptoms such as vertigo, tinnitus, or dizziness. Geometric mean vibratory displacements at threshold ranged from 0.55 nm at 250 Hz to 0.03 nm at 6 kHz. MANOVA demonstrated a repeatability of MVA thresholds. Conclusion: Malleus vibration audiometry will not allow exact linkage of actual implantable hearing aid. But the present study demonstrates that MVA can provide an audiometric tool for assessing ossicular function and integrity prior to implantation of an electronic hearing amplifier.

OriginalspracheEnglisch
ZeitschriftLaryngoscope
Jahrgang114
Ausgabenummer11 I
Seiten (von - bis)2012-2020
Seitenumfang9
ISSN0023-852X
DOIs
PublikationsstatusVeröffentlicht - 11.2004

UN SDGs

Dieser Output leistet einen Beitrag zu folgendem(n) Ziel(en) für nachhaltige Entwicklung

  1. SDG 3 – Gesundheit und Wohlergehen
    SDG 3 – Gesundheit und Wohlergehen

Fingerprint

Untersuchen Sie die Forschungsthemen von „Safe and reliable sound threshold measures with direct vibration of the ossicular chain“. Zusammen bilden sie einen einzigartigen Fingerprint.

Zitieren