Zur Hauptnavigation wechseln Zur Suche wechseln Zum Hauptinhalt wechseln

Resolution Distribution in Single-Sided Magnetic Particle Imaging

T.F. Sattel, T. Knopp, S. Biederer, M. Erbe, K. Lüdtke-Buzug, T.M. Buzug

Abstract

In most coil geometries used for magnetic particle imaging (MPI), the specimen has to be positioned in-between the coil sets for investigation. This poses a size limitation, which is overcome when using a single-sided coil configuration. First measurements documenting the feasibility of this concept have been published. In this contribution, properties of the magnetic field generated by a symmetrical and a single-sided MPI scanner are compared. The spatially dependent resolution distribution of single-sided MPI is investigated and related to the simulated magnetic field properties.