Abstract

By design, the range of beam directions in conventional radiosurgery are constrained to an isocentric array. However, the recent introduction of robotic radiosurgery dramatically increases the flexibility of targeting, and as a consequence, beams need be neither coplanar nor isocentric. Such a nonisocentric design permits a large number of distinct beam directions to be used in one single treatment. These major technical differences provide an opportunity to improve upon the well-established principles for treatment planning used with GammaKnife or LINAC radiosurgery. With this objective in mind, our group has developed over the past decade an inverse planning tool for robotic radiosurgery. This system first computes a set of beam directions, and then during an optimization step, weights each individual beam. Optimization begins with a feasibility query, the answer to which is derived through linear programming. This approach offers the advantage of completeness and avoids local optima. Final beam selection is based on heuristics. In this report we present and evaluate a new strategy for utilizing the advantages of linear programming to improve beam selection. Starting from an initial solution, a heuristically determined set of beams is added to the optimization problem, while beams with zero weight are removed. This process is repeated to sample a set of beams much larger compared with typical optimization. Experimental results indicate that the planning approach efficiently finds acceptable plans and that resampling can further improve its efficiency.

OriginalspracheEnglisch
ZeitschriftMedical Physics
Jahrgang33
Ausgabenummer11
Seiten (von - bis)4005-4011
Seitenumfang7
ISSN0094-2405
DOIs
PublikationsstatusVeröffentlicht - 01.01.2006

UN SDGs

Dieser Output leistet einen Beitrag zu folgendem(n) Ziel(en) für nachhaltige Entwicklung

  1. SDG 3 – Gesundheit und Wohlergehen
    SDG 3 – Gesundheit und Wohlergehen
  2. SDG 9 – Industrie, Innovation und Infrastruktur
    SDG 9 – Industrie, Innovation und Infrastruktur

Fingerprint

Untersuchen Sie die Forschungsthemen von „Resampling: An optimization method for inverse planning in robotic radiosurgery“. Zusammen bilden sie einen einzigartigen Fingerprint.

Zitieren