Zur Hauptnavigation wechseln Zur Suche wechseln Zum Hauptinhalt wechseln

Pulse-oximetric measurement of prilocaine-induced methemoglobinemia in regional anesthesia

Peter Soeding*, Matthias Deppe, Hartmut Gehring

*Korrespondierende/r Autor/-in für diese Arbeit

Abstract

BACKGROUND: The Masimo Radical 7® is a new pulse CO oximeter designed to measure methemoglobin. The device has not been evaluated in a clinical setting. METHODS: In this prospective observational study we compared the arterial methemoglobin levels and the corresponding pulse CO-oximetric values of the Radical 7® in regional anesthesia with prilocaine. RESULTS: We analyzed 360 data pairs with methemoglobin values up to 6.6%. The mean bias and limits (±1.96 SD) of the device were 0.27% (±1.33%). CONCLUSION: We found a high degree of agreement in measurement of methemoglobin between the 2 methods.

OriginalspracheEnglisch
ZeitschriftAnesthesia and Analgesia
Jahrgang111
Ausgabenummer4
Seiten (von - bis)1065-1068
Seitenumfang4
ISSN0003-2999
DOIs
PublikationsstatusVeröffentlicht - 01.01.2010

Fördermittel

Financial support for the work: Peter Soeding and Hartmut Gehring are employees of the University Clinic of Schleswig— Holstein, Campus Luebeck, Luebeck, Germany. Device was provided by Masimo, Inc., Germany.

UN SDGs

Dieser Output leistet einen Beitrag zu folgendem(n) Ziel(en) für nachhaltige Entwicklung

  1. SDG 3 – Gesundheit und Wohlergehen
    SDG 3 – Gesundheit und Wohlergehen

Fingerprint

Untersuchen Sie die Forschungsthemen von „Pulse-oximetric measurement of prilocaine-induced methemoglobinemia in regional anesthesia“. Zusammen bilden sie einen einzigartigen Fingerprint.

Zitieren