Preclinical determination of the best functional position for transcatheter heart valves implanted in rapid deployment bioprostheses

Timm Schäfer, Christian Doose, Buntaro Fujita*, Marc Utzenrath, Sandrine Egron, Christoph Schmitz, Smita Scholtz, Maximilian Kütting, Kavous Hakim-Meibodi, Jochen Börgermann, Jan Gummert, Ulrich Steinseifer, Stephan Ensminger

*Korrespondierende/r Autor/-in für diese Arbeit

Abstract

Aims: The aim of this study was to determine the best functional position of a transcatheter heart valve (THV) implanted as a valve-in-valve (ViV) procedure in small rapid deployment valves (RDV) in an in vitro model. Methods and results: A 21 mm Perceval, Enable or INTUITY RDV was mounted into a pulse duplicator and a 23 mm balloon-expandable or a self-expanding THV was deployed (valve-in-valve) in two different positions. Under physiological hydrodynamic conditions, the performance of the THV was characterised by mean transvalvular pressure gradient (MPG), effective orifice area (EOA) and regurgitation volume (RV). Leaflet kinematics were assessed with high-speed video recordings, and X-ray images were acquired. All THV/RDV combinations met ISO requirements regarding hydrodynamic performance. In most cases, the higher position of the THV performed better than the lower one in terms of a lower MPG and increased EOA. Leaflet motion of the implanted THV was impaired in the lower position. In contrast, regurgitation volumes were relatively small and similar, regardless of the THV position. Conclusions: ViV implantation of a THV in a small RDV yielded satisfactory hydrodynamic results. In most cases, a high implantation position achieved lower MPG, higher EOA and a reduced risk of impaired THV leaflet function. Fluoroscopy images of the best functional ViV positions are presented as a blueprint for patient procedures.

OriginalspracheEnglisch
ZeitschriftEuroIntervention
Jahrgang12
Ausgabenummer14
Seiten (von - bis)1706-1714
Seitenumfang9
ISSN1774-024X
DOIs
PublikationsstatusVeröffentlicht - 02.2017

UN SDGs

Dieser Output leistet einen Beitrag zu folgendem(n) Ziel(en) für nachhaltige Entwicklung

  1. SDG 3 – Gesundheit und Wohlergehen
    SDG 3 – Gesundheit und Wohlergehen