Abstract
In high dose rate brachytherapy, needles are inserted into soft tissue and subsequently radioactive sources are used to deliver a high dose inside the target region. While this approach can achieve a steep dose gradient and oers a focused, organ sparing treatment, it also requires a careful positioning of the needles with respect to the tissue. We have previously proposed to use an optical ber embedded in the needle to detect soft tissue deformation. To validate the approach, we have developed an experimental setup to compare the actual needle motion with the motion estimated via the ber. Our results show a good agreement between actual and estimated motion, indicating that optical deformation detection through the needle is possible.
| Originalsprache | Englisch |
|---|---|
| Seiten | 96-100 |
| Seitenumfang | 5 |
| Publikationsstatus | Veröffentlicht - 2012 |
| Veranstaltung | 2th International MICCAI Workshop on Medical Computer Vision - Nice, Frankreich Dauer: 05.10.2012 → 05.10.2012 |
Tagung, Konferenz, Kongress
| Tagung, Konferenz, Kongress | 2th International MICCAI Workshop on Medical Computer Vision |
|---|---|
| Kurztitel | MCV 2012 |
| Land/Gebiet | Frankreich |
| Ort | Nice |
| Zeitraum | 05.10.12 → 05.10.12 |
UN SDGs
Dieser Output leistet einen Beitrag zu folgendem(n) Ziel(en) für nachhaltige Entwicklung
-
SDG 3 – Gesundheit und Wohlergehen
-
SDG 9 – Industrie, Innovation und Infrastruktur
Fingerprint
Untersuchen Sie die Forschungsthemen von „Phantom validation of optical soft tissue navigation for brachytherapy“. Zusammen bilden sie einen einzigartigen Fingerprint.Zitieren
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver