Abstract
Nowadays, it lacks an open, standardized and dynamic interconnection of medical devices. All existing combinations of medical devices consist of isolated solutions with proprietary interfaces, as no common standards for networking and the exchange of data of medical devices exist. This situation leads to confusing operating rooms and inefficient operations. Thus, new strategies need to be developed for the authorization of dynamically interconnected medical devices. Primarily, those concern of an acquisition and methodical adaption of new requirements and risks resulting from this way of interconnection. The approach is to develop a method for a risk analysis for interconnected medical devices, which is structured modular and consists of a risk assessment of the standalone device and a risk analysis for the interconnection considering the risks involved in the transfer of functions. When interconnecting the medical devices the risk analysis of each of the devices is taken and they are compared by a gap analysis. Through this strategy it will be possible to realize a standard-compliant dynamic interconnection of medical products, which would be advantageous both for clinic operators and producers. This paper presents the current situation of the authorization of combined medical devices and proposes a strategy for the risk management of dynamically interconnected medical devices as a substantial part of the authorization.
| Originalsprache | Englisch |
|---|---|
| Titel | 5th Workshop on Medical Cyber-Physical Systems |
| Redakteure/-innen | Volker Turau, Marta Kwiatkowska, Rahul Mangharam, Christoph Weyer |
| Seitenumfang | 4 |
| Band | 36 |
| Erscheinungsort | Dagstuhl, Germany |
| Herausgeber (Verlag) | Schloss Dagstuhl--Leibniz-Zentrum fuer Informatik |
| Erscheinungsdatum | 01.04.2014 |
| Seiten | 133-136 |
| ISBN (Print) | 978-3-939897-66-8 |
| DOIs | |
| Publikationsstatus | Veröffentlicht - 01.04.2014 |
| Veranstaltung | 5th Medical Cyber-Physical Systems Workshop:Medical Device Interoperability, Safety, and Security Assurance - Berlin, Deutschland Dauer: 14.04.2014 → 14.04.2014 |
UN SDGs
Dieser Output leistet einen Beitrag zu folgendem(n) Ziel(en) für nachhaltige Entwicklung
-
SDG 3 – Gesundheit und Wohlergehen
-
SDG 9 – Industrie, Innovation und Infrastruktur
-
SDG 11 – Nachhaltige Städte und Gemeinschaften
-
SDG 12 – Verantwortungsvoller Konsum und Produktion