Abstract

Retinal laser photocoagulation is an established treatment method for many retinal diseases like macula edema or diabetic retinopathy. The selection of the laser parameters is so far based on post treatment evaluation of the lesion size and strength. Due to local pigment variations in the fundus and individual transmission the same laser parameters often lead to an overtreatment. Optoacoustic allows a non invasive monitoring of the retinal temperature increase during retinal laser irradiation by measuring the temperature dependent pressure amplitudes, which are induced by short probe laser pulses. A 75 ns/ 523 nm Nd:YLF was used as a probe laser at a repetition rate of 1 kHz, and a cw / 532 nm treatment laser for heating. A contact lens was modified with a ring-shaped ultrasonic transducer to detect the pressure waves at the cornea. Temperatures were collected for irradiations leading to soft or invisible lesions. Based on this data the threshold for denaturation was found. By analyzing the initial temperature increase, the further temperature development during irradiation could be predicted. An algorithm was found to calculate the irradiation time, which is needed for a soft lesion formation, from the temperature curve. By this it was possible to provide a real-time dosimetry by automatically switching off the treatment laser after the calculated irradiation time. Automatically controlled coagulations appear softer and more uniformly.

OriginalspracheEnglisch
TitelOphthalmic Technologies XXI
Redakteure/-innenFabrice Manns, Per G. Söderberg, Arthur Ho
Seitenumfang8
Band7885
Herausgeber (Verlag)SPIE
Erscheinungsdatum16.02.2011
Aufsatznummer78850T
ISBN (Print)978-081948422-2
DOIs
PublikationsstatusVeröffentlicht - 16.02.2011
VeranstaltungOphthalmic Technologies XXI - San Francisco, USA / Vereinigte Staaten
Dauer: 22.01.201124.01.2011
Konferenznummer: 84404

Strategische Forschungsbereiche und Zentren

  • Forschungsschwerpunkt: Biomedizintechnik

Fingerprint

Untersuchen Sie die Forschungsthemen von „Optoacoustic temperature determination and automatic coagulation control in rabbits“. Zusammen bilden sie einen einzigartigen Fingerprint.

Zitieren