Abstract
OBJECTIVES: In this study, results of a functional in vitro study of 2 newly developed valved stents for transcatheter mitral valve implantation are presented. METHODS: Two novel stent designs, an oval-shaped and a D-shaped stent with a strut fixation system were developed. The fixation force of the novel stents were tested in vitro in porcine hearts with a tensile test set-up. In further experiments, the stents were equipped with a circular valved stent, and the valve performances were investigated in a pulsatile heart valve tester. RESULTS: Sufficient mean stent fixation forces in the range of 24.2 ± 0.9N to 28.6 ± 1.9N were measured for the different stent models. The novel valved stents showed good performance in an in vitro pulsatile heart valve tester. A sufficient opening area and low opening pressures were measured for all tested mitral valved stents. Compared with an established reference valve, the D-shaped stent and the oval-shaped valved stent showed a lower systolic transvalvular pressure gradient, which indicates slightly greater extent of valvular leakage of the closed valved stents. However, the mitral nitinol valved stents demonstrated adequate durability. CONCLUSIONS: This study indicates a sufficient annular fixation force of the tested transcatheter mitral valve implantation valved stent prototypes. Therefore, these mitral valved stents demonstrate a new type of mitral valved stent design.
| Originalsprache | Englisch |
|---|---|
| Zeitschrift | Interactive Cardiovascular and Thoracic Surgery |
| Jahrgang | 26 |
| Ausgabenummer | 2 |
| Seiten (von - bis) | 190-195 |
| Seitenumfang | 6 |
| ISSN | 1569-9293 |
| DOIs | |
| Publikationsstatus | Veröffentlicht - 01.02.2018 |
Fördermittel
This work was supported by the German Center of Cardiovascular Research (DZHK).
UN SDGs
Dieser Output leistet einen Beitrag zu folgendem(n) Ziel(en) für nachhaltige Entwicklung
-
SDG 3 – Gesundheit und Wohlergehen
Fingerprint
Untersuchen Sie die Forschungsthemen von „Novel stent design for transcatheter mitral valve implantation“. Zusammen bilden sie einen einzigartigen Fingerprint.Zitieren
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver