Modelling of Scatter in the System Matrix for 3D PET Image Reconstruction: A Comparative Study

Moritz Schaar, Magdalena Rafecas

Abstract

Compton scattering in positron emission tomography (PET) is a major image degrading effect. In this work, an analytical model for Compton scatter is included in the system matrix of two iterative image reconstruction techniques, maximum likelihood expectation maximization (MLEM) and origin ensemble (OE); their performance with respect to conventional MLEM and MLEM with analytic single scatter simulation (SSS) based scatter correction is validated according to the NEMA NU 2-2007 protocol. Background variability is reduced by 9.4% on average compared to MLEM with SSS. Contrast recovery coefficients (CRCs) are improved for larger sources.

OriginalspracheEnglisch
Titel2019 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)
Herausgeber (Verlag)IEEE
Erscheinungsdatum10.2019
Aufsatznummer9059797
ISBN (Print)978-1-7281-4165-7
ISBN (elektronisch)978-1-7281-4164-0
DOIs
PublikationsstatusVeröffentlicht - 10.2019
Veranstaltung2019 IEEE Nuclear Science Symposium and Medical Imaging Conference - Manchester, Großbritannien / Vereinigtes Königreich
Dauer: 26.10.201902.11.2019
Konferenznummer: 159071

Strategische Forschungsbereiche und Zentren

  • Forschungsschwerpunkt: Biomedizintechnik

DFG-Fachsystematik

  • 2.22-32 Medizinische Physik, Biomedizinische Technik

Fingerprint

Untersuchen Sie die Forschungsthemen von „Modelling of Scatter in the System Matrix for 3D PET Image Reconstruction: A Comparative Study“. Zusammen bilden sie einen einzigartigen Fingerprint.

Zitieren