Megahertz intravascular Doppler optical coherence tomography enables simultaneous morphological and flow pattern imaging

Tianshi Wang*, Tom Pfeiffer, Joost Daemen, Frits Mastik, Wolfgang Wieser, A. F.W. Van Der Steen, Robert Huber, Gijs Van Soest

*Korrespondierende/r Autor/-in für diese Arbeit

Abstract

We demonstrate megahertz intravascular Doppler Optical Coherence Tomography (OCT). The OCT system relies on a 1.1 mm diameter motorized catheter and a 1.5 MHz Fourier Domain Mode Locked (FDML) laser. By resolving the phase shift between adjoint A-lines, the flow pattern image can be reconstructed. Imaging experiments were carried out in swine coronary artery in vitro at a speed of 600 frames/s. The MHz sweep rate not only allows us to investigate flow velocity of up to 37.5 cm/s without phase-unwrapping, but also enables dynamic flow imaging at high frame rate.

OriginalspracheEnglisch
TitelDiagnostic and Therapeutic Applications of Light in Cardiology 2019
Redakteure/-innenKenton W. Gregory , Laura Marcu
Seitenumfang5
Band10855
Herausgeber (Verlag)SPIE
Erscheinungsdatum26.02.2019
Aufsatznummer1085503
DOIs
PublikationsstatusVeröffentlicht - 26.02.2019
VeranstaltungDiagnostic and Therapeutic Applications of Light in Cardiology 2019 - San Francisco, USA / Vereinigte Staaten
Dauer: 02.02.201903.02.2019
Konferenznummer: 148274

Strategische Forschungsbereiche und Zentren

  • Forschungsschwerpunkt: Biomedizintechnik

Fingerprint