Abstract
We demonstrate megahertz intravascular Doppler Optical Coherence Tomography (OCT). The OCT system relies on a 1.1 mm diameter motorized catheter and a 1.5 MHz Fourier Domain Mode Locked (FDML) laser. By resolving the phase shift between adjoint A-lines, the flow pattern image can be reconstructed. Imaging experiments were carried out in swine coronary artery in vitro at a speed of 600 frames/s. The MHz sweep rate not only allows us to investigate flow velocity of up to 37.5 cm/s without phase-unwrapping, but also enables dynamic flow imaging at high frame rate.
Originalsprache | Englisch |
---|---|
Titel | Diagnostic and Therapeutic Applications of Light in Cardiology 2019 |
Redakteure/-innen | Kenton W. Gregory , Laura Marcu |
Seitenumfang | 5 |
Band | 10855 |
Herausgeber (Verlag) | SPIE |
Erscheinungsdatum | 26.02.2019 |
Aufsatznummer | 1085503 |
DOIs | |
Publikationsstatus | Veröffentlicht - 26.02.2019 |
Veranstaltung | Diagnostic and Therapeutic Applications of Light in Cardiology 2019 - San Francisco, USA / Vereinigte Staaten Dauer: 02.02.2019 → 03.02.2019 Konferenznummer: 148274 |
Strategische Forschungsbereiche und Zentren
- Forschungsschwerpunkt: Biomedizintechnik