Abstract
The goal in breast conserving surgery (BCS) of ductal carcinoma in situ (DCIS) is removal of the tumor with a clear surgical margin. However, re-excision rates are regrettably high. To date, there are no adequate procedures for intraoperative margin assessment of DCIS. A multicenter, single arm study was conducted to evaluate the benefit of a novel device (MarginProbe®) in intraoperative margin assessment during BCS of DCIS, the associated reduction of re-excisions and the cosmetic outcome of the treated patients. We present results of 42 patients enrolled in 3 German institutions. The device was used as an adjunctive tool to standard of care. The device use was associated with a reduction in re-excision rates by 56%, from 39% to 17% (p=0.018).
Originalsprache | Englisch |
---|---|
Zeitschrift | Breast |
Jahrgang | 23 |
Ausgabenummer | 1 |
Seiten (von - bis) | 94-96 |
Seitenumfang | 3 |
ISSN | 0960-9776 |
DOIs | |
Publikationsstatus | Veröffentlicht - 02.2014 |