Magnetic Particle Imaging: Artifact-Free Metallic Stent Lumen Imaging in a Phantom Study

2 Zitate (Scopus)

Abstract

Purpose: To illustrate the potential of magnetic particle imaging (MPI) for stent lumen imaging in comparison with clinical computed tomography (CT) and magnetic resonance imaging (MRI). Materials and Methods: Imaging of eight tracer-filled, stented vessel phantoms and a tracer-filled, non-stented reference phantom for each diameter was performed on a preclinical MPI scanner: eight commercially available coronary stents of different dimensions (diameter: 3–4 mm; length: 11–38 mm) and materials (stainless steel, platinum–chromium) were implanted into silicone vessel phantoms. For comparison, all vessel phantoms were also visualized by MRI and CT. Two radiologists assessed the images regarding stent-induced artifacts using a 5-point grading scale. Results: The visualization of all stented vessel phantoms was achieved without stent-induced artifacts with MPI. In contrast, MRI and CT images revealed multiform stent-induced artifacts. Conclusion: Given its clinical introduction, MPI has the potential to overcome the disadvantages of MRI and CT concerning the visualization of the stent lumen.

OriginalspracheEnglisch
ZeitschriftCardioVascular and Interventional Radiology
Jahrgang43
Ausgabenummer2
Seiten (von - bis)331-338
Seitenumfang8
ISSN0174-1551
DOIs
PublikationsstatusVeröffentlicht - 01.02.2020