Abstract
Linear optical coherence tomography (LOCT) systems are a simple and robust alternative to time-domain optical coherence tomography systems, but a detector with approximately 104 pixels is needed for an imaging depth of 2 mm. We present a new system for LOCT with a special mask attached to the image sensor. The mask essentially performs a downconversion of the spatial frequencies by multiplication with a second spatial frequency. This reduces the fringe frequency of the optical coherence tomography signal so that the signal can be sampled with fewer pixels.
Originalsprache | Englisch |
---|---|
Zeitschrift | Optics Letters |
Jahrgang | 29 |
Ausgabenummer | 14 |
Seiten (von - bis) | 1644-1646 |
Seitenumfang | 3 |
ISSN | 0146-9592 |
DOIs | |
Publikationsstatus | Veröffentlicht - 15.07.2004 |
Strategische Forschungsbereiche und Zentren
- Forschungsschwerpunkt: Biomedizintechnik