Large-area dynamic contrast MHz optical coherence tomography for label-free imaging of porcine tissue

Sazgar Burhan, Madita Göb, Mario Pieper, Tjalfe Laedtke, Thorge Grahl, Michael Münter, Hinnerk Schulz-Hildebrandt, Gereon Hüttmann, Peter König, Robert Huber

Abstract

We demonstrate a 3.2 MHz-OCT system for inter-volumetric dynamic optical coherence tomography of ex vivo porcine kidney tissue. Employing a home-built Fourier Domain mode locking (FDML) laser with a 1310 nm wavelength, the system achieved a lateral resolution of 3.48 μm and a frame rate of 612 Hz. A motorized XYZ positioning stage enabled the precise acquisition of multiple volumes, which were seamlessly stitched together to generate a comprehensive dataset with a total area of 2.6 × 2.6 mm2. Validations against histological sections confirmed the system’s ability to visualize cellular tissue structures.
OriginalspracheEnglisch
TitelOptical Coherence Tomography and Coherence Domain Optical Methods in Biomedicine XXIX
Redakteure/-innenRainer A. Leitgeb, Yoshiaki Yasuno
Seitenumfang1
Band13305
Herausgeber (Verlag)SPIE
Erscheinungsdatum03.2025
Seiten1330502
DOIs
PublikationsstatusVeröffentlicht - 03.2025

Fingerprint

Untersuchen Sie die Forschungsthemen von „Large-area dynamic contrast MHz optical coherence tomography for label-free imaging of porcine tissue“. Zusammen bilden sie einen einzigartigen Fingerprint.

Zitieren