Abstract
We demonstrate a 3.2 MHz-OCT system for inter-volumetric dynamic optical coherence tomography of ex vivo porcine kidney tissue. Employing a home-built Fourier Domain mode locking (FDML) laser with a 1310 nm wavelength, the system achieved a lateral resolution of 3.48 μm and a frame rate of 612 Hz. A motorized XYZ positioning stage enabled the precise acquisition of multiple volumes, which were seamlessly stitched together to generate a comprehensive dataset with a total area of 2.6 × 2.6 mm2. Validations against histological sections confirmed the system’s ability to visualize cellular tissue structures.
Originalsprache | Englisch |
---|---|
Titel | Optical Coherence Tomography and Coherence Domain Optical Methods in Biomedicine XXIX |
Redakteure/-innen | Rainer A. Leitgeb, Yoshiaki Yasuno |
Seitenumfang | 1 |
Band | 13305 |
Herausgeber (Verlag) | SPIE |
Erscheinungsdatum | 03.2025 |
Seiten | 1330502 |
DOIs | |
Publikationsstatus | Veröffentlicht - 03.2025 |