Joint aperture detection for speckle reduction and increased collection efficiency in ophthalmic MHz OCT

Thomas Klein, Raphael André, Wolfgang Wieser, Tom Pfeiffer, Robert Huber*

*Korrespondierende/r Autor/-in für diese Arbeit
40 Zitate (Scopus)

Abstract

Joint-aperture optical coherence tomography (JA-OCT) is an angle-resolved OCT method, in which illumination from an active channel is simultaneously probed by several passive channels. JA-OCT increases the collection efficiency and effective sensitivity of the OCT system without increasing the power on the sample. Additionally, JA-OCT provides angular scattering information about the sample in a single acquisition, so the OCT imaging speed is not reduced. Thus, JA-OCT is especially suitable for ultra high speed in-vivo imaging. JA-OCT is compared to other angle-resolved techniques, and the relation between joint aperture imaging, adaptive optics, coherent and incoherent compounding is discussed. We present angleresolved imaging of the human retina at an axial scan rate of 1.68 MHz, and demonstrate the benefits of JA-OCT: Speckle reduction, signal increase and suppression of specular and parasitic reflections. Moreover, in the future JA-OCT may allow for the reconstruction of the full Doppler vector and tissue discrimination by analysis of the angular scattering dependence.

OriginalspracheEnglisch
ZeitschriftBiomedical Optics Express
Jahrgang4
Ausgabenummer4
Seiten (von - bis)619-634
Seitenumfang16
DOIs
PublikationsstatusVeröffentlicht - 01.04.2013

Fingerprint

Untersuchen Sie die Forschungsthemen von „Joint aperture detection for speckle reduction and increased collection efficiency in ophthalmic MHz OCT“. Zusammen bilden sie einen einzigartigen Fingerprint.

Zitieren