Abstract
PURPOSE: To report three cases with breakage of a 20-gauge kryptonite forceps during vitreoretinal surgery. DESIGN: Interventional case report. METHODS: Pars plana vitrectomy that included a membrane peeling was performed in three patients through standard ports with 20-gauge vitrectomy systems. The peeling procedure was attempted with the end-gripping kryptonite forceps. RESULTS: Intraoperatively, the branch/tip broke and dropped onto the posterior pole. Removal of the broken part was uncomplicated, and the subsequent clinical course was unremarkable. In one case, the handle and branch/tip part were exchanged inadvertently. In the other two cases, neither an exchange nor a maladaptation of the parts was noted. CONCLUSION: An exchange of the hand piece and gripping end may increase the risk of breakage and retinal damage. The delicate gripping tips, however, may break because of inadequate maintenance and/or aging of the material. Intraoperative visual and functional examination is recommended before intraocular use.
| Titel in Übersetzung | Intraoperativer Bruch einer 20-Gauge-Tano-Pinzette |
|---|---|
| Originalsprache | Englisch |
| Zeitschrift | American Journal of Ophthalmology |
| Jahrgang | 141 |
| Ausgabenummer | 1 |
| Seiten (von - bis) | 215-217 |
| Seitenumfang | 3 |
| ISSN | 0002-9394 |
| DOIs | |
| Publikationsstatus | Veröffentlicht - 01.01.2006 |
UN SDGs
Dieser Output leistet einen Beitrag zu folgendem(n) Ziel(en) für nachhaltige Entwicklung
-
SDG 3 – Gesundheit und Wohlergehen
Fingerprint
Untersuchen Sie die Forschungsthemen von „Intraoperativer Bruch einer 20-Gauge-Tano-Pinzette“. Zusammen bilden sie einen einzigartigen Fingerprint.Zitieren
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver