Abstract
Differential signaling can provide a higher fidelity of the receive signals in MPI, which is especially important for mobile scanner setups not being operated in a shielding room. A passive, highly symmetrical filter is presented as a part of the fully differential receive chain of the scanner setup. Due to the combined use with a gradiometric receive coil, the requirements for the components in terms of voltage and current specifications can be lowered which allows for partly building the filter on a PCB with well-matched SMD capacitors and trace lengths. The achieved attenuation for the differential-mode excitation field feedthrough is 77 dB while providing a common-mode rejection of more than 49 dB throughout the receive bandwidth.
Originalsprache | Englisch |
---|---|
Aufsatznummer | 2009031 |
Zeitschrift | International Journal on Magnetic Particle Imaging |
Jahrgang | 6 |
Ausgabenummer | 2 |
Seiten (von - bis) | 1-3 |
Seitenumfang | 3 |
ISSN | 2365-9033 |
DOIs | |