Abstract
We present full volumetric high speed OCT imaging of the retina with multiple settings varying in volume size and volume rate. The volume size ranges from 255x255 A-scans to 160x40 A-scans with 450 samples per depth scan with volume rates varying between 20.8 V/s for the largest volumes to 195.2 V/s for the smallest. The system is based on a 1060nm Fourier domain mode locked (FDML) laser with 1.6MHz line rate. Scanning along the fast axis is performed with a 2.7 kHz or 4.3 kHz resonant scanner operated in bidirectional scanning mode, while a standard galvo scanner is used for the slow axis. The performance is analyzed with respect to various potential applications, like intraoperative OCT.
Originalsprache | Englisch |
---|---|
Titel | Optical Coherence Imaging Techniques and Imaging in Scattering Media |
Redakteure/-innen | Brett E. Bouma, Maciej Wojtkowski |
Seitenumfang | 8 |
Band | 95410Z |
Herausgeber (Verlag) | SPIE |
Erscheinungsdatum | 22.07.2015 |
Seiten | 95410Z-95410Z-8 |
ISBN (Print) | 9781628417067 |
DOIs | |
Publikationsstatus | Veröffentlicht - 22.07.2015 |
Veranstaltung | EUROPEAN CONFERENCES ON BIOMEDICAL OPTICS - München, Deutschland Dauer: 21.06.2015 → 25.06.2015 Konferenznummer: 131538 https://spie.org/conferences-and-exhibitions?webSyncID=8ea9e3e7-7b6b-4836-788d-177c8547e3b2&sessionGUID=7a4397d1-7ae8-5844-19e9-c62bcc516bb5&_ga=2.220047066.216408110.1526970521-1856364488.1526970521&SSO=1 |