Zur Hauptnavigation wechseln Zur Suche wechseln Zum Hauptinhalt wechseln

Hammerstein–Wiener Motion Artifact Correction for Functional Near-Infrared Spectroscopy: A Novel Inertial Measurement Unit-Based Technique

Hayder R. Al-Omairi, Arkan AL-Zubaidi, Sebastian Fudickar, Andreas Hein, Jochem W. Rieger*

*Korrespondierende/r Autor/-in für diese Arbeit

Abstract

Participant movement is a major source of artifacts in functional near-infrared spectroscopy (fNIRS) experiments. Mitigating the impact of motion artifacts (MAs) is crucial to estimate brain activity robustly. Here, we suggest and evaluate a novel application of the nonlinear Hammerstein–Wiener model to estimate and mitigate MAs in fNIRS signals from direct-movement recordings through IMU sensors mounted on the participant’s head (head-IMU) and the fNIRS probe (probe-IMU). To this end, we analyzed the hemodynamic responses of single-channel oxyhemoglobin (HbO) and deoxyhemoglobin (HbR) signals from 17 participants who performed a hand tapping task with different levels of concurrent head movement. Additionally, the tapping task was performed without head movements to estimate the ground-truth brain activation. We compared the performance of our novel approach with the probe-IMU and head-IMU to eight established methods (PCA, tPCA, spline, spline Savitzky–Golay, wavelet, CBSI, RLOESS, and WCBSI) on four quality metrics: SNR, △AUC, RMSE, and R. Our proposed nonlinear Hammerstein–Wiener method achieved the best SNR increase (p < 0.001) among all methods. Visual inspection revealed that our approach mitigated MA contaminations that other techniques could not remove effectively. MA correction quality was comparable with head- and probe-IMUs.

OriginalspracheEnglisch
Aufsatznummer3173
ZeitschriftSensors
Jahrgang24
Ausgabenummer10
ISSN1424-8220
DOIs
PublikationsstatusVeröffentlicht - 05.2024

Fördermittel

H.R.A.-O. was supported by the University of Technology—Iraq fellowship development program. J.W.R. was funded by a DFG Device Center Grant INST 184/216-1“Tools and infrastructure for open and reproducible, neuroimaging” and DFG-grant RI 1511/3-1.

TrägerTrägernummer
Technische Universiteit Delft
Deutsche ForschungsgemeinschaftINST 184/216-1

    UN SDGs

    Dieser Output leistet einen Beitrag zu folgendem(n) Ziel(en) für nachhaltige Entwicklung

    1. SDG 3 – Gesundheit und Wohlergehen
      SDG 3 – Gesundheit und Wohlergehen
    2. SDG 9 – Industrie, Innovation und Infrastruktur
      SDG 9 – Industrie, Innovation und Infrastruktur
    3. SDG 10 – Weniger Ungleichheiten
      SDG 10 – Weniger Ungleichheiten

    Fingerprint

    Untersuchen Sie die Forschungsthemen von „Hammerstein–Wiener Motion Artifact Correction for Functional Near-Infrared Spectroscopy: A Novel Inertial Measurement Unit-Based Technique“. Zusammen bilden sie einen einzigartigen Fingerprint.