Abstract
Full volumetric high speed OCT imaging of the retina with multiple settings varying in volume size and volume rate is presented. The volume size ranges from 255x255 A-scans to 160x40 A-scans with 450 samples per depth scan. The volume rates vary between 20.8 V/s for the largest volumes to 195.2 V/s for the smallest. The system is based on a 1060nm Fourier domain mode locked (FDML) laser with 1.6MHz line rate. Scanning along the fast axis is performed with a 2.7 kHz or 4.3 kHz resonant scanner operated in bidirectional scanning mode, while a standard galvo scanner is used for the slow axis. The performance is analyzed with respect to various potential applications, like intraoperative OCT.
| Originalsprache | Englisch |
|---|---|
| Titel | Optical Coherence Tomography and Coherence Domain Optical Methods in Biomedicine XIX |
| Redakteure/-innen | V. V. Tuchin, James G. Fujimoto, Joseph A. Izatt |
| Seitenumfang | 7 |
| Band | 931202 |
| Herausgeber (Verlag) | SPIE |
| Erscheinungsdatum | 16.03.2015 |
| Seiten | 931202-931202-7 |
| ISBN (Print) | 9781628414028 |
| DOIs | |
| Publikationsstatus | Veröffentlicht - 16.03.2015 |
| Veranstaltung | SPIE BIOS - San Francisco, USA / Vereinigte Staaten Dauer: 07.02.2015 → 12.02.2015 |
UN SDGs
Dieser Output leistet einen Beitrag zu folgendem(n) Ziel(en) für nachhaltige Entwicklung
-
SDG 9 – Industrie, Innovation und Infrastruktur
Fingerprint
Untersuchen Sie die Forschungsthemen von „Full volumetric video rate OCT of the posterior eye with up to 195.2 volumes/s“. Zusammen bilden sie einen einzigartigen Fingerprint.Zitieren
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver