Zur Hauptnavigation wechseln Zur Suche wechseln Zum Hauptinhalt wechseln

Five-year clinical follow-up of a randomized comparison of a polymer-free sirolimus-eluting stent versus a polymer-based paclitaxel-eluting stent in patients with diabetes mellitus (LIPSIA Yukon trial)

Thomas Stiermaier*, Anja Heinz, Denis Schloma, Klaus Kleinertz, Wilfried Dänschel, Sandra Erbs, Axel Linke, Enno Boudriot, Bernward Lauer, Gerhard Schuler, Holger Thiele, Steffen Desch

*Korrespondierende/r Autor/-in für diese Arbeit

Abstract

Objectives The long-term performance of polymer-free stent systems in patients with diabetes mellitus has not been investigated extensively. This study reports long-term results of the LIPSIA Yukon trial which compared the polymer-free sirolimus-eluting Yukon Choice stent with the polymer-based paclitaxel-eluting Taxus Liberté stent in this subpopulation. At 9 months, the Yukon Choice stent failed to show non-inferiority in terms of the primary end point late lumen loss, while no significant difference in clinical outcome was detected. Methods and Results The LIPSIA Yukon trial randomized 240 patients with diabetes mellitus to a polymer-free sirolimus eluting stent (Yukon Choice, Translumina) versus a polymer-based paclitaxel-eluting stent (Taxus Liberté, Boston Scientific). Clinical follow-up was conducted with a standardized telephone follow-up and all events were centrally adjudicated. Follow-up was available for 98.3% of patients after a median of 5.0 years. The incidence of all-cause death (16.9% versus 14.0%, P = 0.67), respectively definite or presumed cardiovascular death (7.6% versus 8.8%, P = 0.94) were similar in the Yukon Choice and the Taxus Liberté group. There were no significant differences in the rates of myocardial infarction (9.3% versus 7.9%, P = 0.88), definite stent thrombosis (0.8% versus 0.9%, P = 1.0), target lesion revascularization (15.3% versus 15.8%, P = 1.0), target vessel revascularization (18.6% versus 23.7%, P = 0.44), non-target vessel revascularization (18.6% versus 26.3%, P = 0.21), and stroke (3.4% versus 4.4%, P = 0.96) between patients assigned to the Yukon Choice and the Taxus Liberté stent. Conclusion At 5 years of follow-up, clinical outcome was similar between the polymer-free sirolimus-eluting Yukon Choice stent and the polymer-based paclitaxel-eluting Taxus Liberté stent. © 2013 Wiley Periodicals, Inc.

OriginalspracheEnglisch
ZeitschriftCatheterization and Cardiovascular Interventions
Jahrgang83
Ausgabenummer3
Seiten (von - bis)418-424
Seitenumfang7
ISSN1522-1946
DOIs
PublikationsstatusVeröffentlicht - 15.02.2014

UN SDGs

Dieser Output leistet einen Beitrag zu folgendem(n) Ziel(en) für nachhaltige Entwicklung

  1. SDG 3 – Gesundheit und Wohlergehen
    SDG 3 – Gesundheit und Wohlergehen

Strategische Forschungsbereiche und Zentren

  • Forschungsschwerpunkt: Gehirn, Hormone, Verhalten - Center for Brain, Behavior and Metabolism (CBBM)

Fingerprint

Untersuchen Sie die Forschungsthemen von „Five-year clinical follow-up of a randomized comparison of a polymer-free sirolimus-eluting stent versus a polymer-based paclitaxel-eluting stent in patients with diabetes mellitus (LIPSIA Yukon trial)“. Zusammen bilden sie einen einzigartigen Fingerprint.

Zitieren