Fabrication and Evaluation of 3D-Printed 89Zr Image Quality Phantoms for Preclinical Immuno-PET Imaging

  • Weill Cornell Medical College
  • Department of Nuclear Medicine and Clinical Molecular Imaging, University Hospital Tuebingen
  • Department of Preclinical Imaging and Radiopharmacy, Werner Siemens Imaging Center, Eberhard Karls University Tuebingen

Abstract

Introduction: Immuno-PET relies on labelling monoclonal antibodies with isotopes with sufficient long half-life times, such as 89Zr. Compared with 18F, 89Zr has a lower positron yield resulting in increased image noise, particularly at late imaging time points; multi-day 89Zr imaging is further complicated by challenges in phantom preparation, as 89Zr solutions are prone to adsorption to phantom walls, tubing, and syringes, leading to unstable and non-uniform activity distributions. These effects limit the reproducibility and accuracy of performance and protocol optimization studies. To address these challenges, we investigate the use of stereolithographic 3D printing of 89Zr-containing structures, enabling wall-less, stable, and reproducible activity distributions for multi-day immuno-PET phantom studies.

Methods: 3D printing was performed using an LCD-based Photon Mono printer (Anycubic, China). The printer operates by selectively curing a liquid photopolymer resin using a UV-emitting LCD screen and provides nominal resolutions of 51 × 51 × 10 µm in the X, Y, and Z directions, respectively. A modified resin–radionuclide mixing protocol was developed by combining 89Zr with a water-washable photopolymer resin (density 1.1 g/cm³). Moreover, non-radioactive zirconium diluted in hydrochloric acid was added to promote binding between the resin and 89Zr. The mixture with 0.2 MBq/mL was homogenized by vigorous manual shaking without external heating. A single-point phantom was fabricated by depositing a high-activity resin droplet directly above the LCD screen at predefined locations and was compared with four commercial 22Na point sources to assess spatial precision. CT imaging was used to assess the geometric characteristics of the sources. Additionally, two phantoms, design based on NEMA NU 4-2008 IQ phantoms, were printed with (89Zr 0.2 MBq/mL) and with (18F, 1.8 MBq/mL) and compared to a filled phantom with (89Zr 0.2 MBq/mL).

Results: In the CT image of the commercial point sources, three radioactive seeds were identified within a 20 mm diameter area, each with an approximate diameter of 0.8 mm and all skewed toward one side of the disk. This asymmetry implies that, for PET systems with spatial resolution < 1 mm, rotating the source by 180° could lead to different spatial resolution measurements due to altered seed positioning. In contrast, the 3D-printed point source consisted of a single, centrally located seed, providing accurate reproducibility regardless of the source orientation. For the IQ phantoms, uniformity measurements following NEMA NU 4-2008 showed mean activity concentrations of 1.7 MBq/mL (%STD = 4.6) for the 18F printed phantom, 0.2 MBq/mL (%STD = 3.4) for the 89Zr printed phantom, and 0.2 MBq/mL (%STD = 1.5) for the 89Zr filled phantom, indicating comparable values between the 89Zr printed and conventionally filled phantoms. Recovery coefficients for the printed phantoms were similar to those of the filled phantom, as illustrated in Figure 2.

Conclusions: This study demonstrates the feasibility of manufacturing 3D-printed image quality phantoms incorporating 89Zr for preclinical PET imaging. The single-point source phantom achieved accurate centering of a 1 mm source seed, eliminating geometric ambiguities observed in commercial multi-seed point sources. While such discrepancies are negligible in clinical PET systems, they may influence measurements in high-resolution preclinical or dedicated scanners. For example, when modeling the point spread function (PSF) for image reconstruction, precise source positioning becomes advantageous. Using a modified mixing protocol, an 89Zr-based IQ phantom was successfully printed and showed comparable performance to conventional fillable phantoms and 18F-printed phantoms. This approach enables stable, reproducible phantom designs suitable for multi-day immuno-PET studies. Future work will focus on translation to clinical scanners and evaluation of multi-day imaging performance.
OriginalspracheEnglisch
PublikationsstatusVeröffentlicht - 05.2026

UN SDGs

Dieser Output leistet einen Beitrag zu folgendem(n) Ziel(en) für nachhaltige Entwicklung

  1. SDG 3 – Gesundheit und Wohlergehen
    SDG 3 – Gesundheit und Wohlergehen
  2. SDG 9 – Industrie, Innovation und Infrastruktur
    SDG 9 – Industrie, Innovation und Infrastruktur

Strategische Forschungsbereiche und Zentren

  • Forschungsschwerpunkt: Biomedizintechnik
  • Zentren: Center of Imaging Sciences (CIS)

DFG-Fachsystematik

  • 2.22-32 Medizinische Physik, Biomedizinische Technik

Fingerprint

Untersuchen Sie die Forschungsthemen von „Fabrication and Evaluation of 3D-Printed 89Zr Image Quality Phantoms for Preclinical Immuno-PET Imaging“. Zusammen bilden sie einen einzigartigen Fingerprint.

Dieses zitieren