Abstract
System matrix and x-space resonstruction are two of the main approaches for image reconstruction of field free line magnetic particle imaging. A comparative study of both options is performed on data from a phantom measurement in a permanent magnet-based scanner system. The system matrix reconstruction is performed in a hybrid fashion with data obtained in a spectrometer. This data is also used to obtain the relaxation time and particle diameter of the particles used. A deconvolution can then be used to enhance the image quality of the x-space approach which is compared to the system matrix reconstruction. A slight blurring can be depicted in the x-space reconstruction but overall a good agreement between both approaches is reached with the Structural Similarity Index yielding a value of 0.79.
Originalsprache | Englisch |
---|---|
Aufsatznummer | 2203061 |
Zeitschrift | International Journal on Magnetic Particle Imaging |
Jahrgang | 8 |
Ausgabenummer | 1 |
ISSN | 2365-9033 |
DOIs | |
Publikationsstatus | Veröffentlicht - 2022 |