Abstract
Purpose: To evaluate the feasibility of 2D-perfusion angiography (2D-PA) for detecting leakage of the double-balloon catheter used for chemosaturation percutaneous hepatic perfusion (CS-PHP). Materials and Methods: Overall, 112 CS-PHP (09/2015-09/2018) in 52 patients were retrospectively screened for leakage alongside the double-balloon catheter on standard venograms. Finally, 18 procedures with visually detected leakage were included. Fifteen consecutive procedures without leakage served as control. To evaluate 2D-PA for leakage detection, the acquired digital subtraction venograms were post-processed. For each balloon, two different target ROIs were evaluated to assess a possible impact of localization and shape of the ROIs. Time to peak (TTP), peak density (PD), area under the curve (AUC), and ratios of target ROI/reference ROIs (PDtROI/PDREF; AUCtROI/AUCREF; and TTPtROI/TTPREF) were calculated. Results: Leakages were located as follows: 15/18 cranial and 3/18 caudal. At the cranial balloon both ROIs showed a significant decrease in PDtROI/PDREF and AUCtROI/AUCREF (ROI1: p < 0.0001; p < 0.0001; ROI2: p < 0.0001; p < 0.0001) and a significant increase in TTPtROI/TTPREF (ROI1: p = 0.0009; ROI2: p = 0.0003) after double-balloon correction. Following balloon adjustment, the 2D-PA ratios (PD and AUC) of the tested ROIs differed significantly (p < 0.05). The inter-individual comparison of the 2D-PA parameters of the group with leakage before balloon correction and the non-leakage group showed significantly different 2D-PA values for the cranial balloon in both ROIs (p < 0.05). No significant differences were found for the caudal balloon. Conclusion: 2D-PA provides a feasible tool for detecting leakages alongside the cranial portion of the double-balloon catheter used in CS-PHP. The shape and position of the ROIs used to assess perfusion and flow have an impact on the measurements.
| Originalsprache | Englisch |
|---|---|
| Zeitschrift | CardioVascular and Interventional Radiology |
| Jahrgang | 42 |
| Ausgabenummer | 10 |
| Seiten (von - bis) | 1441-1448 |
| Seitenumfang | 8 |
| ISSN | 0174-1551 |
| DOIs | |
| Publikationsstatus | Veröffentlicht - 15.10.2019 |
UN SDGs
Dieser Output leistet einen Beitrag zu folgendem(n) Ziel(en) für nachhaltige Entwicklung
-
SDG 3 – Gesundheit und Wohlergehen
Fingerprint
Untersuchen Sie die Forschungsthemen von „Chemosaturation Percutaneous Hepatic Perfusion (CS-PHP) with Melphalan: Evaluation of 2D-Perfusion Angiography (2D-PA) for Leakage Detection of the Venous Double-Balloon Catheter“. Zusammen bilden sie einen einzigartigen Fingerprint.Zitieren
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver