T1 - Chemosaturation Percutaneous Hepatic Perfusion (CS-PHP) with Melphalan: Evaluation of 2D-Perfusion Angiography (2D-PA) for Leakage Detection of the Venous Double-Balloon Catheter
AU - Dewald, Cornelia L.A.
AU - Meine, Timo C.
AU - Winther, Hinrich M.B.
AU - Kloeckner, Roman
AU - Maschke, Sabine K.
AU - Kirstein, Martha M.
AU - Vogel, Arndt
AU - Wacker, Frank K.
AU - Meyer, Bernhard C.
AU - Renne, Julius
AU - Hinrichs, Jan B.
N1 - Publisher Copyright:
© 2019, Springer Science+Business Media, LLC, part of Springer Nature and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE).
PY - 2019/10/15
Y1 - 2019/10/15
N2 - Purpose: To evaluate the feasibility of 2D-perfusion angiography (2D-PA) for detecting leakage of the double-balloon catheter used for chemosaturation percutaneous hepatic perfusion (CS-PHP). Materials and Methods: Overall, 112 CS-PHP (09/2015-09/2018) in 52 patients were retrospectively screened for leakage alongside the double-balloon catheter on standard venograms. Finally, 18 procedures with visually detected leakage were included. Fifteen consecutive procedures without leakage served as control. To evaluate 2D-PA for leakage detection, the acquired digital subtraction venograms were post-processed. For each balloon, two different target ROIs were evaluated to assess a possible impact of localization and shape of the ROIs. Time to peak (TTP), peak density (PD), area under the curve (AUC), and ratios of target ROI/reference ROIs (PDtROI/PDREF; AUCtROI/AUCREF; and TTPtROI/TTPREF) were calculated. Results: Leakages were located as follows: 15/18 cranial and 3/18 caudal. At the cranial balloon both ROIs showed a significant decrease in PDtROI/PDREF and AUCtROI/AUCREF (ROI1: p < 0.0001; p < 0.0001; ROI2: p < 0.0001; p < 0.0001) and a significant increase in TTPtROI/TTPREF (ROI1: p = 0.0009; ROI2: p = 0.0003) after double-balloon correction. Following balloon adjustment, the 2D-PA ratios (PD and AUC) of the tested ROIs differed significantly (p < 0.05). The inter-individual comparison of the 2D-PA parameters of the group with leakage before balloon correction and the non-leakage group showed significantly different 2D-PA values for the cranial balloon in both ROIs (p < 0.05). No significant differences were found for the caudal balloon. Conclusion: 2D-PA provides a feasible tool for detecting leakages alongside the cranial portion of the double-balloon catheter used in CS-PHP. The shape and position of the ROIs used to assess perfusion and flow have an impact on the measurements.
AB - Purpose: To evaluate the feasibility of 2D-perfusion angiography (2D-PA) for detecting leakage of the double-balloon catheter used for chemosaturation percutaneous hepatic perfusion (CS-PHP). Materials and Methods: Overall, 112 CS-PHP (09/2015-09/2018) in 52 patients were retrospectively screened for leakage alongside the double-balloon catheter on standard venograms. Finally, 18 procedures with visually detected leakage were included. Fifteen consecutive procedures without leakage served as control. To evaluate 2D-PA for leakage detection, the acquired digital subtraction venograms were post-processed. For each balloon, two different target ROIs were evaluated to assess a possible impact of localization and shape of the ROIs. Time to peak (TTP), peak density (PD), area under the curve (AUC), and ratios of target ROI/reference ROIs (PDtROI/PDREF; AUCtROI/AUCREF; and TTPtROI/TTPREF) were calculated. Results: Leakages were located as follows: 15/18 cranial and 3/18 caudal. At the cranial balloon both ROIs showed a significant decrease in PDtROI/PDREF and AUCtROI/AUCREF (ROI1: p < 0.0001; p < 0.0001; ROI2: p < 0.0001; p < 0.0001) and a significant increase in TTPtROI/TTPREF (ROI1: p = 0.0009; ROI2: p = 0.0003) after double-balloon correction. Following balloon adjustment, the 2D-PA ratios (PD and AUC) of the tested ROIs differed significantly (p < 0.05). The inter-individual comparison of the 2D-PA parameters of the group with leakage before balloon correction and the non-leakage group showed significantly different 2D-PA values for the cranial balloon in both ROIs (p < 0.05). No significant differences were found for the caudal balloon. Conclusion: 2D-PA provides a feasible tool for detecting leakages alongside the cranial portion of the double-balloon catheter used in CS-PHP. The shape and position of the ROIs used to assess perfusion and flow have an impact on the measurements.
UR - http://www.scopus.com/inward/record.url?scp=85065968050&partnerID=8YFLogxK
U2 - 10.1007/s00270-019-02243-4
DO - 10.1007/s00270-019-02243-4
M3 - Journal articles
C2 - 31089781
AN - SCOPUS:85065968050
SN - 0174-1551
VL - 42
SP - 1441
EP - 1448
JO - CardioVascular and Interventional Radiology
JF - CardioVascular and Interventional Radiology
IS - 10
ER -