Biomechanical performance of dental implants inserted in different mandible locations and at different angles: A finite element study

Barbora Thomková, Petr Marcián*, Libor Borák, Marek Joukal, Jan Wolff

*Korrespondierende/r Autor/-in für diese Arbeit

Abstract

Statement of problem: Accurate implant placement is essential for the success of dental implants. This placement influences osseointegration and occlusal forces. The freehand technique, despite its cost-effectiveness and time efficiency, may result in significant angular deviations compared with guided implantation, but the effect of angular deviations on the stress-strain state of peri-implant bone is unclear. Purpose: The purpose of this finite element analysis (FEA) study was to examine the effects of angular deviations on stress-strain states in peri-implant bone. Material and methods: Computational modeling was used to investigate 4 different configurations of dental implant positions, each with 3 angles of insertion. The model was developed using computed tomography images, and typical mastication forces were considered. Strains were analyzed using the mechanostat hypothesis. Results: The location of the implant had a significant impact on bone strain intensity. An angular deviation of ±5 degrees from the planned inclination did not significantly affect cancellous bone strains, which primarily support the implant. However, it had a substantial effect on strains in the cortical bone near the implant. Such deviations also significantly influenced implant stresses, especially when the support from the cortical bone was uneven or poorly localized. Conclusions: In extreme situations, angular deviations can lead to overstraining the cortical bone, risking implant failure from unfavorable interaction with the implant. Accurate implant placement is essential to mitigate these risks.

OriginalspracheEnglisch
ZeitschriftJournal of Prosthetic Dentistry
Jahrgang131
Ausgabenummer1
Seiten (von - bis)128.e1-128.e10
ISSN0022-3913
DOIs
PublikationsstatusVeröffentlicht - 01.2024

Fördermittel

Funding Supported by the Brno University of Technology [project FSI-S-23–8240], Brno, Czech Republic.

TrägerTrägernummer
Vysoké Učení Technické v BrněFSI-S-23–8240

    UN SDGs

    Dieser Output leistet einen Beitrag zu folgendem(n) Ziel(en) für nachhaltige Entwicklung

    1. SDG 3 – Gesundheit und Wohlergehen
      SDG 3 – Gesundheit und Wohlergehen
    2. SDG 9 – Industrie, Innovation und Infrastruktur
      SDG 9 – Industrie, Innovation und Infrastruktur

    Strategische Forschungsbereiche und Zentren

    • Forschungsschwerpunkt: Biomedizintechnik

    DFG-Fachsystematik

    • 2.22-28 Zahnheilkunde; Mund-, Kiefer- und Gesichtschirurgie
    • 2.22-32 Medizinische Physik, Biomedizinische Technik

    Fingerprint

    Untersuchen Sie die Forschungsthemen von „Biomechanical performance of dental implants inserted in different mandible locations and at different angles: A finite element study“. Zusammen bilden sie einen einzigartigen Fingerprint.

    Zitieren